Evaluation of the cobas 4800 CT/NG test for detecting Chlamydia trachomatis and Neisseria gonorrhoeae

Abstract

Objectives To investigate the performance of the fully automated cobas 4800 CT/NG test for detection of Chlamydia trachomatis and Neisseria gonorrhoeae.

Methods The study was conducted using 900 clinical specimens (496 urine and 404 swab specimens) for C trachomatis testing, of which 498 specimens (318 urine and 180 swab specimens) were also tested for N gonorrhoeae. The results of the cobas 4800 CT/NG test were compared with those obtained from the Roche COBAS AMPLICOR CT/NG and COBAS TaqMan CT assays. N gonorrhoeae-positive specimens were further tested using in-house, real-time PCR assays. A panel of 223 Neisseria isolates was used to further investigate the performance of the cobas 4800 N gonorrhoeae assay.

Results For urine specimens, the sensitivity, specificity and negative and positive predictive values of the cobas 4800 CT/NG test were 94.5%, 99.5%, 98.8% and 97.7%, respectively, for C trachomatis, and 92.9%, 100%, 99.7% and 100%, respectively, for N gonorrhoeae. For swab specimens, the sensitivity, specificity and negative and positive predictive values were 92.0%, 100%, 99.5% and 100%, respectively, for C trachomatis, and 100%, 99.4%, 100% and 90.0%, respectively, for N gonorrhoeae. All N gonorrhoeae isolates were positive and all non-gonococcal Neisseria strains were negative by the cobas 4800 N gonorrhoeae assay.

Conclusions The cobas 4800 CT/NG test is suitable for high through-put identification of C trachomatis and N gonorrhoeae infections.

  • Chlamydia trachomatis
  • DNA amplification
  • Neisseria gonorrhoea

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.