Table 1

Repeat LCR testing and PCR testing of initially positive LCR urines during the Wirral Chlamydia Pilot (Sept 1999 to Oct 2000, baseline) and for 3 month periods since the issue of the device correction (February 2001)

No of urinesPCR+PCR+/−PCR−PCR (a)PCR (b)
(a) Inhibitory, (b) insufficient.
Initial LCR positive (Sep 99–Oct 00)960
    Repeat LCR:
        Positive883 (92%)*****Not done****
        Equivocal (0.5–0.99) 12 (1.3%) 6 6
        Negative 65 (6.8%)13502
Initial LCR positive (Mar–May 01)134
    Repeat LCR:
        Positive 74 (55%)701 3
        Equivocal (0.5–0.99) 18 515
        Negative 42 (31%) 636
Initial LCR positive (Jun–Aug 01)121
    Repeat LCR:
        Positive 95 (79%)903 2
        Equivocal (0.5–0.99) 2 (1.7%) 2
        Negative 24 (19.8%) 519
Initial LCR positive (Sep–Nov 01)90
    Repeat LCR:
        Positive 87 (96.6%)823 11
        Equivocal (0.5–0.99) 1 (1.1%) 1
        Negative 2 (2.2%) 2