Evaluation of a novel electrochemical detection method for Chlamydia trachomatis: application for point-of-care diagnostics

IEEE Trans Biomed Eng. 2011 Mar;58(3):755-8. doi: 10.1109/TBME.2010.2095851. Epub 2010 Nov 29.

Abstract

Atlas Genetics has developed a point-of-care device for Chlamydia trachomatis utilizing a novel electrochemical detection principle. The assay has a time-to-result of less than 25 min. An independent preclinical validation study using 306 pretyped clinical samples determined a clinical sensitivity of 98.1% and specificity of 98.0%.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Chlamydia Infections / diagnosis
  • Chlamydia Infections / microbiology
  • Chlamydia trachomatis / genetics
  • Chlamydia trachomatis / isolation & purification*
  • Electrochemical Techniques / instrumentation*
  • Humans
  • Microfluidic Analytical Techniques / instrumentation*
  • Point-of-Care Systems*
  • Polymerase Chain Reaction / instrumentation
  • Reproducibility of Results
  • Sensitivity and Specificity