Table 2

Diagnostic accuracy characteristics for rapid tests for Chlamydia trachomatis, Neisseria gonorrhoeae and syphilis, compared with PCR and rapid plasma reagin (RPR) and Treponema pallidum haemagglutination assay (TPHA) tests for syphilis

POC testNTPFNFPTNSENSItivity (95% CI)SPECIFICIty (95% CI)LR (+)LR (−)
N. gonorrhoeae
Acon for NG7730821744NQ*97.2 (96 to 98.5)NQNQ
Acon Duo49117148212.5 (0 to 41.7)99.8 (99.3 to 100)60.40.9
C. Trachomatis
Acon for CT229517020722.7 (2.9 to 42.5)100 (99.7 to 100)NQ0.8
Acon Duo4911841143130.5 (17.9 to 43.1)99.8 (99.2 to 100)131.80.7
Quick view6642033460737.74 (23.7 to 51.7)99.4 (98.6 to 100)57.60.6
Syphilis
Acon1439120171410100 (95.8 to 100)98.8 (98.2 to 99.4)84.0NQ
SD Bioline140011117137191.6 (71.8 to 100)98.8 (98.2 to 99.4)74.30.1
  • *Not quantifiable.

  • FN, false negative; FP, false positive; LR, likelihood ratio; N, number; NQ, not quantifiable; POC, point of care; TN, true negative; TP, true positive.